Abstract

Titanium carbide coatings have a broad range of biomedical applications because of their high hardness, low friction, excellent corrosion resistance, and good biocompatibility. NiTi alloys are also widely used in surgical implants in orthodontics and orthopedics. In order to improve the surface properties, nanostructured titanium carbide coatings are deposited on NiTi by plasma immersion ion implantation and deposition after a titanium interlayer has been fabricated on the NiTi substrate. The structure and corrosion behavior which impact the biological properties are investigated systematically by X-ray diffraction, X-ray photoelectron spectroscopy, field-emission scanning electron microscopy, atomic force microscopy, and electrochemical impedance spectroscopy in simulated body fluids at 37°C. The TiC thin films with a C/Ti ratio of 1.087 have the (220) orientation. The EIS results demonstrate that the Ti/TiC multilayer provides significantly better corrosion resistance and stability compared to the uncoated NiTi substrate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.