Abstract

Background: This finite element analysis (FEA) evaluated stress distribution in implant-supported overdenture (ISO) and peri-implant bone using one extracoronal (ball) and two intracoronal (locator and Zest Anchor Advanced Generation (ZAAG)) attachment systems. Methods: In this in vitro study, the mandible was modelled in the form of an arc-shaped bone block with 33 mm height and 8 mm width. Two titanium implants were modelled at the site of canine teeth, and three attachments (ZAGG, locator, and ball) were placed over them. Next, 100 N load was applied at 90° and 30° angles from the molar site of each quadrant to the implants. The stress distribution pattern in the implants and the surrounding bone was analyzed, and the von Mises stress around the implants and in the crestal bone was calculated. Results: While minimum stress in peri-implant bone following load application at 30° angle was noted in the mesial point of the locator attachment, maximum stress was recorded at the distal point of the ball attachment following load application at 90° angle. Maximum stress around the implant following load application at 90° angle was noted in the lingual point of the ball attachment while minimum stress was recorded in the lingual point of the locator attachment following load application at 90° angle. Conclusions: According to the results, the locator attachment is preferred to the ZAAG attachment, and the ball attachment should be avoided if possible.

Highlights

  • This finite element analysis (FEA) evaluated stress distribution in implant-supported overdenture (ISO) and peri-implant bone using one extracoronal and two intracoronal (locator and Zest Anchor Advanced Generation (ZAAG)) attachment systems

  • Comparison of the von Mises stress in the peri-implant bone in use of the three attachment types revealed maximum stress in the ball attachment (6598) followed by the ZAAG (5455) and locator (4760) attachments after load application at 90° angle

  • Maximum stress was noted in the ball attachment (6254) followed by the ZAAG (5330) and locator (4551) attachments after load application at 30° angle

Read more

Summary

Introduction

This finite element analysis (FEA) evaluated stress distribution in implant-supported overdenture (ISO) and peri-implant bone using one extracoronal (ball) and two intracoronal (locator and Zest Anchor Advanced Generation (ZAAG)) attachment systems. Results: While minimum stress in peri-implant bone following load application at 30° angle was noted in the mesial point of the locator attachment, maximum stress was recorded at the distal point of the ball attachment following load application at 90° angle. Maximum stress around the implant following load application at 90° angle was noted in the lingual point of the ball attachment while minimum stress was recorded in the lingual point of the locator attachment following load application at 90° angle. Implant-supported overdenture (ISO) is an efficient treatment plan due to its simplicity, non-invasiveness, and affordability [1,2]. It is suitable for patients who are not satisfied with their complete dentures. Very low stress can lead to disuse atrophy of bone while

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.