Abstract
ABSTRACTObjective: In this study, simulations were performed by the finite element method (FEM) to determine the tension and displacement in mini-implants and in expander appliance during rapid maxillary expansion, by varying the number and location of the mini-implants. Methods: For the computational simulation, a three-dimensional mesh was used for the maxilla, mini-implants and expander appliance. Comparisons were made on six different Mini-implant Assisted Rapid Palatal Expander (MARPE) configurations, by varying the amount and location of mini-implants. A closed suture was design and received two activations of 0.25 mm, and an open suture had a 0.5-mm aperture that received 20 activations, also of 0.25 mm. Results: For the closed suture, the maximum displacement values in the mini-implants were between 0.253 and 0.280 mm, and stress was between 1,348.9 and 2,948.2 MPa; in the expander appliance, the displacement values were between 0.256 and 0.281 mm, and stress was between 738.52 and 1,207.6 MPa. For the open suture, the maximum displacement values in the mini-implants were between 2.57 and 2.79 mm, and stress was between 5,765.3 and 10,366 MPa; in the appliance, the maximum displacements was between 2.53 and 2.89 mm, and stress was between 4,859.7 and 9,157.4 MPa. Conclusions: There were higher stress concentrations in the mini-implant than in the expander arm. In the simulations with a configuration of three mini-implants, stress overload was observed in the isolated mini-implant. Displacements of the mini-implants and arms of the appliance were similar in all simulations.
Highlights
Mini-implant Assisted Rapid Palatal Expander (MARPE) is an appliance for correction of maxillary atresia in adults as an alternative to surgical procedures.[1,2,3,4,5] The mechanical behaviour of the mini-implants and expander appliance during maxillary disjunction is important, especially due to the heavy forces applied to perform the procedure.[6]
In this study, simulations were performed by the finite element method (FEM) to determine the tension and displacement in mini-implants and in expander appliance during rapid maxillary expansion, by varying the number and location of the mini-implants
For the computational simulation, a three-dimensional mesh was used for the maxilla, mini-implants and expander appliance
Summary
Mini-implant Assisted Rapid Palatal Expander (MARPE) is an appliance for correction of maxillary atresia in adults as an alternative to surgical procedures.[1,2,3,4,5] The mechanical behaviour of the mini-implants and expander appliance during maxillary disjunction is important, especially due to the heavy forces applied to perform the procedure.[6] The number of mini-implants required for MARPE varies according to the technique and the clinical indication: appliance configurations with two or four mini-implants are more commonly observed.[1,2,3,4,7,8]. The finite element method (FEM) is a valuable resource for investigating orthodontic mechanics.[9,10,11] Stress and strain simulations using the FEM have been shown to be useful for improving the MARPE behavior. Hybrid expanders with two mini-implants for anchorage,[9] the effects on the nasomaxillary complex[11] and, recently, a comparison between mono- and bicortical anchorage using the MARPE12 have been examined
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.