Abstract
This work involved two major parts: the first one is the experimental part which included treatment of scoliosis deformity by manufacturing thoracolumbosacral orthosis, measuring the cobb angle of deformity, measuring the gait cycle data and walk path for both legs and suggesting a composite material to improve the mechanical properties of the orthosis and finally the interface pressure between trunk and orthosis is measured for twelve points covering of the total TLSO surface area by using f-socket devise. The second part of this study is the numerical simulation part during which the stresses are calculated using Ansys software for calculating stresses due to interface pressure loading boundary condition. The result shows no deference in gait cycle phases but the clear difference noted in walking path due to deviate center of mass, maximum pressure recorded left thoracic region with 900KPa due to correct spinal deformity while the minimum pressure recorded at right chest with l40KPa because of it is tissue region and Maximum value of stress was recorded at the left thoracic region with 2.81MPa due to Maximum interface pressure at this point.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.