Abstract

Burch-Schneider cages are often used for the treatment of acetabular bone defects. In several clinical studies these cages have shown good mid- to long-term results. However, a higher failure rate has been reported in large Paprosky IIIB defects compared with smaller Paprosky II-IIIA defects. This study aims to investigate the effect of cage support on cage failure by means of finite element analysis. The Von Mises stresses in both the implant and the bone are analyzed for a Burch-Schneider cage used in the following scenarios: (1) a large acetabular bone defect, (2) a small acetabular bone defect and (3) a large acetabular bone defect in combination with a reinforcement plate. The results show that implant and bone stresses are higher in the large defect (99th percentile of 146.6 and 73.5 MPa respectively) than in the small defect (99th percentile of 43.9 and 47.9 MPa respectively). Adding a reinforcement plate to posteriorly support the cage decreases the stresses but not fully compensates for the missing bone support (99th percentile of 93.1 and 55.3 MPa respectively). Since high stresses cause an increased risk for fatigue failure and implant loosening, sufficient implant support is required to reduce the risk of cage failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.