Abstract
Purpose: To determine the variation in strength of clear corneal incisions, as demonstrated by degree of incision leakage when challenged by increased intraocular pressure, in relationship to the incision width and length.Setting: Mackool Eye Institute, Astoria, New York.Methods: Clear corneal incisions 3.0 or 3.5 mm in width and from 1.0 to 3.5 mm in length (0.5 mm increments) were studied in human cadaver whole globes. Pressure was applied at the corneal apex or 8.0 mm posterior to the external wound margin to determine the strength of the incision with the application of external force.Results: Clear corneal incisions of 3.0 or 3.5 mm in width and at least 2.0 mm in length demonstrated substantially greater resistance to incision failure than shorter incision lengths with both apical and posterior applied forces.Conclusion: Clear corneal incisions 2.0 mm or greater in length demonstrate resistance to leakage comparable to similarly constructed scleral tunnel incisions.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.