Abstract

The Breakthrough Devices Program of the US Food and Drug Administration has accelerated the development and evaluation of medical devices for patients with heart failure. One such device is the Optimizer Smart System, which the US Food and Drug Administration approved in 2019. The Optimizer device was evaluated in a pivotal randomized clinical trial (FIX-HF-5C [Confirmatory Randomized Trial Evaluating the Optimizer System]) that leveraged Bayesian borrowing of external data to reduce the sample size and determine therapeutic device benefit versus continued medical therapy. Bayesian borrowing is explained in the context of the FIX-HF-5C trial, including an overview of the statistical methodologies, regulatory considerations, and interpretations of trial results. The US Food and Drug Administration Breakthrough Devices Program and novel Bayesian statistical methodology accelerated the path to regulatory approval and patient access to a potentially lifesaving device and may serve as a model for future clinical trials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.