Abstract

PurposeTo develop a method for mapping the B1 transmit (B1t) and B1 receive (B1r) fields from two gradient echo datasets each with a different flip angle and from these two images obtain accurate T1 and proton density (PD) maps of the brain. MethodsA strategically acquired gradient echo (STAGE) data set is collected using two flip angles each with multiple echoes. The B1t field extraction was based on forcing cortical gray matter and white matter to have specific T1 values and fitting the resulting B1t field to a quadratic function. The B1r field extraction was based on synthesizing isointense images despite there being two or three tissue types present in the brain. This method was tested on 10 healthy volunteers and 20 stroke patients from data acquired at 3.0Tesla. ResultsWith the knowledge of the B1t and B1r fields, the uniformity of tissue T1 and PD maps was considerably improved. T1 values were measured for both the midbrain and basal ganglia and found to be in good agreement with the literature. Discussion and ConclusionsSTAGE provides a practical way to assess the B1t and the B1r fields which can then be used to correct for spatial variations in the images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.