Abstract

The purpose of this study was to assess the potential for optimization of mechanical ventilator waveforms using multiple frequencies of oscillatory flow delivered simultaneously to minimize the risk of ventilator-induced lung injury (VILI) associated with regional strain, strain rate, and mechanical power. Optimization was performed using simulations of distributed oscillatory flow and gas transport in a computational model of anatomically derived branching airway segments and viscoelastic terminal acini under healthy and injured conditions. Objective functions defined by regional strain or strain rate were minimized by single-frequency ventilation waveforms using the highest or lowest frequencies available, respectively. However, a mechanical power objective function was minimized by a combination of multiple frequencies delivered simultaneously. This simulation study thus demonstrates the potential for multifrequency oscillatory ventilation to reduce regional mechanical power in comparison to single-frequency ventilation, and thereby reduce the risk of VILI.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.