Abstract

ObjectivesTo assess whether strain histograms are equal to strain ratios in predicting breast tumour malignancy and to see if either could be used to upgrade Breast Imaging Reporting and Data System (BI-RADS) 3 tumours for immediate biopsy.MethodsNinety-nine breast tumours were examined using B-mode BI-RADS scorings and strain elastography. Strain histograms and ratios were assessed, and areas- under-the-receiver-operating-characteristic-curve (AUROC) for each method calculated. In BI-RADS 3 tumours cut-offs for strain histogram and ratio values were calculated to see if some tumours could be upgraded for immediate biopsy. Linear regression was performed to evaluate the effect of tumour depth and size, and breast density on strain elastography.ResultsForty-four of 99 (44.4%) tumours were malignant. AUROC of BI-RADS, strain histograms and strain ratios were 0.949, 0.830 and 0.794 respectively. There was no significant difference between AUROCs of strain histograms and strain ratios (P = 0.405), while they were both inferior to BI-RADS scoring (P<0.001, P = 0.008). Four out of 26 BI-RADS 3 tumours were malignant. When cut-offs of 189 for strain histograms and 1.44 for strain ratios were used to upgrade BI-RADS 3 tumours, AUROCS were 0.961 (Strain histograms and BI-RADS) and 0.941 (Strain ratios and BI-RADS). None of them was significantly different from BI-RADS scoring alone (P = 0.249 and P = 0.414). Tumour size and depth, and breast density influenced neither strain histograms (P = 0.196, P = 0.115 and P = 0.321) nor strain ratios (P = 0.411, P = 0.596 and P = 0.321)ConclusionStrain histogram analyses are reliable and easy to do in breast cancer diagnosis and perform comparably to strain ratio analyses. No significant difference in AUROCs between BI-RADS scoring and elastography combined with BI-RADS scoring was found in this study.

Highlights

  • Elastography has emerged as an auspicious adjunct to B-mode ultrasound examination [1,2,3]

  • There was no significant difference between areas- under-the-receiveroperating-characteristic-curve (AUROC) of strain histograms and strain ratios (P = 0.405), while they were both inferior to Breast Imaging Reporting and Data System (BI-RADS) scoring (P

  • None of them was significantly different from BI-RADS scoring alone (P = 0.249 and P = 0.414)

Read more

Summary

Introduction

Elastography has emerged as an auspicious adjunct to B-mode ultrasound examination [1,2,3]. Tumour stiffness is correlated with the risk of malignancy [1, 2, 5] Both quantitative (shear-wave elastography) and qualitative methods (strain elastography) have been introduced [5]. The current study concerns strain elastography, in which the relative depiction of tissue stiffness makes visual classifications of tumour strain or semi quantifications necessary. The ratio is calculated as the ratio between the strain in reference tissue ROIs (Region Of Interest) and the strain in breast tumour ROIs—the higher the strain ratio, the stiffer the tumour [7]. Tumour stiffness is calculated as mean pixel colour values of a single ROI within the tumour [10]. Strain histograms are independent of reference ROIs, and are easy to perform, even if the surrounding tissue is inhomogeneous. Most authors agree that elastography should be used mainly for reassessments of tumours with high risks of either false negative or false positive results, such as in BI-RADS 3 and 4a tumours [11,12]

Objectives
Methods
Results
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.