Abstract
In connection with comparative measurements on dosimeters used in therapy at various European betatrons the distribution was measured also of the unwanted bremsstrahlung and of fast neutrons. As bremsstrahlung detector use was made of a portable cylindrical ionization chamber. The results show that the distribution of the unwanted bremsstrahlung depends on the construction of the betatron. The measurement of the fast neutrons produced by nuclear photoeffects was carried out on a paraffin double moderator by means of a BF3 counter. In this way it was possible to determine either the neutron dose rate or flux density and so to establish the mean neutron energy. The distribution of the fast neutrons is determined by the shape and dimensions of the therapy room. The measurements thus provide important clues for a suitable construction of therapy rooms. (auth)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: RöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.