Abstract
Spencer-Attix stopping power ratios for the dosimetry of high-energy photon beams used in radiation therapy have been calculated using the Monte Carlo method. The stopping power ratios are calculated in a more consistent way than previously and are given as a function of the attenuation properties of the beam. The dependence of the stopping power ratio on the electron contamination of the beam as well as on depth and field size has also been investigated. Results are compared with stopping power ratios recommended in different dosimetry protocols and to experimental results. The agreement with most dosimetry protocols is within about one per cent and with recent experimental data is better than half a per cent.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.