Abstract
A new method for performing comprehensive GC x GC in the stop-flow mode is presented. A device was used to pneumatically stop the flow in the first dimension ((1)D) (by applying pressure pulses at the junction between the two columns), while flow was maintained in the second dimension ((2)D). This allowed for better preservation of resolution in the (1)D of the GC x GC chromatograms, and the extension of the (2)D's separation space, reducing or eliminating the extent of wraparound. When increased flow rates in the (2)D were used, sensitivity enhancements were also observed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.