Abstract
We report in this paper an instrument for measuring the Stokes parameters of a light beam. This Stokes polarimeter employs two low birefringence photoelastic modulators (PEMs) operating at different resonant frequencies. The two PEMs in the instrument are oriented 45° apart. An analyzer, following the PEM-pair, bisects the PEMs' optical axes. Electronic signals from a detector are analyzed using lock-in amplifiers at the PEMs' first and second harmonics. A computer program calculates and displays the intensity parameter and the normalized Stokes parameters of the light beam measured. Common laboratory lasers are measured as examples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.