Abstract
We calculate the photometric Stokes parameter covariance matrices and SNRs estimated by polarimeters exposed to general noise distributions, such as mixed Poisson-Gaussian (PG) noise. The measurement model includes the effects of optical losses and detector quantum efficiency, enabling quantitative comparison of instruments that have different photometric efficiencies. We demonstrate this capability by comparing the performance of many common polarimeter configurations, including diattenuator-based systems, such as Azzam's four-detector polarimeter [Opt. Lett.10, 309 (1985)OPLEDP0146-959210.1364/OL.10.000309] and Kudenov's stacked photovoltaic polarimeter [Opt. Express24, 14737 (2016)OPEXFF1094-408710.1364/OE.24.014737]. Working with the full covariance matrix under mixed PG noise, we also show that instruments optimized under assumptions of Gaussian noise simultaneously exhibit optimal behavior under Poisson noise.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.