Abstract

The use of cobalt chromium (CoCr) in orthopaedic joint replacement shields the peri-implant bone stress, contributing to a premature loosening of the implants. In order to reduce the need for revision surgeries, light weight implants with tailored functionalities need to be developed. In this study, the compressive mechanical properties of laser-melted CoCr cellular structures with a pillar octahedral architecture [0°±45°] were investigated. Four types of graded cellular structures, based on grading orientations along radial and longitudinal planes, were manufactured using selective laser melting techniques. The cellular structures in this study have the mechanical properties (E=2.3–3.1GPa, σ=113–523MPa) compatible with bone structures. Grading a porosity of the CoCr cellular structures provides a greater stress transfer to the proximal peri-implant area. The axially graded cellular structures demonstrated significant reduction of the peri-implant stress shielding. Incorporation of CoCr graded cellular structures into a structure like femoral stems is expected to have the potential to reduce the revision surgeries.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call