Abstract
Reconstruction for a large area of sternal defect is challenging. The available sternal reconstruction techniques and materials have their limitations. We designed a modular sternal reconstruction system made of titanium alloy. This system was applied to a patient who received partial sternal resection for well-differentiated chondrosarcoma of the sternum. The sternal neoplasm measured 80×70 mm2. After resection of most of the sternal body and adjacent ribs, standard modules of adequate types and sizes were chosen and assembled during operation. The operation lasted for 240 min, with an intraoperative blood loss of 100 mL. The patient was followed up for 32 months, while the prosthesis remained stable, with no signs of infection or looseness.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.