Abstract

A 54-year-old male with partial denture underwent stereotactic radiosurgery with an infrared camera-guided system for a metastatic brain tumor arising from lung cancer. Although this method utilizes a biteplate mounted on the upper jaw to detect head movement, the patient only had four teeth in his upper jaw. In order to stabilize the biteplate, the maxillary denture was fixed to the biteplate with an autopolymerizing resin. In addition, the rest-occlusal position of the lower jaw was impressed on the inferior surface of the biteplate with an autopolymerizing resin. To assess reproducibility and stability, the distance between the left and right incus and left and right markers was measured during pre-planning, as well as before and after stereotactic irradiation. Wearing the biteplate ensures the accuracy of radiotherapy planning for the implementation of radiosurgery in patients who have many maxillary teeth missing. However, a large degree of error was observed when the biteplate was removed.