Abstract

The present study comparatively evaluates the impact of energy-matched flattening filter-free (FFF) photon beams with different energy levels on the physical-dosimetric quality of lung and liver stereotactic body radiotherapy (SBRT) treatment plans. For this purpose, 54different lung and liver lesions from 44patients who had already received SBRT combined with volumetric modulated arc therapy (VMAT) were included in this retrospective planning study. Planning computed tomography scans already available were used for the renewed planning with 6MV, 6MV-FFF, 10MV, and 10MV-FFF under constant planning objectives. The treatment delivery data, dosimetric distributions, and dose-volume histograms as well as parameters such as the conformity index and gradient indices were the basis for the evaluation and comparison of treatment plans. Asignificant reduction of beam-on time (BOT) was achieved due to the high dose rates of FFF beams. In addition, we showed that for FFF beams compared to flattened beams of the same energy level, smaller planning target volumes (PTV) require fewer monitor units (MU) than larger PTVs. An equal to slightly superior target volume coverage and sparing of healthy tissue as well as organs at risk in both lung and liver lesions were found. Significant differences were seen mainly in the medium to lower dose range. We found that FFF beams together with VMAT represent an excellent combination for SBRT of lung or liver lesions with shortest BOT for 10MV-FFF but significant dose savings for 6MV-FFF in lung lesions.