Abstract

The surface of polymethacrylate monoliths was functionalized by a post-polymerization modification, by means of a novel photo-initiated graft procedure where the charged monomer, sulfopropyl methacrylate, was controllably grafted stepwise, i.e. with incremental graft energies. The grafting approach was optimized using scanning capacitively coupled contactless conductivity detection. The effect of the localized ion exchange capacity and resultant gradient stationary phase upon ion-exchange chromatographic retention, selectivity, and performance was investigated, and compared to a homogeneously grafted (isotropic) column. The gradient column provided reduced peak widths at half height for both cationic analytes, with a reduction of 34 and 33%, respectively, when compared to the isotropic column.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.