Abstract

The single-device simplicity for mechanical thrombectomy (MT) is now challenged by the complementary efficacy of dual-device first-line with a stent retriever and an aspiration catheter. To compare the outcomes after MT initiated with a single device vs dual devices in acute anterior circulation large vessel occlusion. Patients who underwent MT for acute internal carotid artery (ICA) or M1 occlusion between 2015 and 2020 were retrospectively analyzed. We divided patients into 2 groups: single-device first-line, defined as patients who underwent first-device pass with either a stent retriever or aspiration catheter, and dual-device first-line, defined as first-device pass with both devices. One hundred forty-one patients were in the single-device group, and 119 were in the dual-device group. In the dual-device group, coiling or kinking of the extracranial ICA was more frequent ( P = .07) and the guide catheters were less frequently navigated to the ICA ( P < .001). 37% of the single-device group was converted to dual-device use. The proportions of mTICI ≥ 2c after the first pass were similar (33% vs 32%. adjusted odds ratio 0.91, 95% CI 0.51-1.62). An mRS score of 0 to 2 at 3 months was achieved similarly (53% vs 48%, P = .46). The total cost for thrombectomy devices was lower in the single-device group ( P < .001). The proportions of first-pass mTICI ≥ 2c were not different between the 2 groups with similar functional outcomes, although the dual-device group more likely to have unfavorable vascular conditions.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call