Abstract

PurposeTo evaluate the safety and long-term clinical outcomes of stent-graft placement to treat hepatic arterial hemorrhageafter pancreaticobiliary surgery. Materials and MethodsOutcomes were retrospectively evaluated in 61 patients (50 men and 11 women; mean age, 63 years) who underwent stent-graft placement for delayed arterial hemorrhage (after 24 hours) after pancreaticobiliary surgery from 2006 to 2023. Bleeding sites included the gastroduodenal artery stump (n = 54), common or proper hepatic artery (n = 5), and right hepatic artery (n = 2). The stent-grafts used were Viabahn (n = 27), Comvi (n = 11), Jostent (n = 3), Covera (n = 11), and Lifestream (n = 7). Technical and clinical success and adverse events (AE) were evaluated. After stent-graft placement, overall survival (OS), hemorrhage-free survival (HFS), and stent patency were evaluated. ResultsThe technical and clinical success rates of stent-graft placement were 97% and 93%, respectively. The severe AE rate was 12% and was significantly higher in patients who underwent pylorus-sacrificing rather than pylorus-preserving surgery (P = .001). None of the severe AEs were associated with patient mortality. Median OS after stent-graft placement was 854 days, and median HFS was 822 days. The 1-, 3-, 5-, and 10-year stent patency rates were 87%, 84%, 79%, and 72%, respectively. ConclusionsStent-graft placement was safe and provided long-term control of hepatic arterial hemorrhage after pancreaticobiliary surgery.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call