Abstract

Endovascular treatment of a giant extracranial internal carotid aneurysm by a stent graft implantation was unsuccessful due to a high flow leak directly through the stent graft's coating. The problem was solved deploying a second stent graft inside the previously implanted one resulting in complete exclusion of the aneurysmal sac and patent carotid lumen preservation. The review of the literature did not provide a case using this endovascular strategy. Follow-up for more than 12 months, using CT angiography, showed confirmed aneurysmal exclusion and carotid patency and no clinical complications have been detected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.