Abstract

Peripheral neural interface (PNI) with a stable integration of synthetic elements with neural tissue is key for successfulneuro-prosthetic applications. An inevitable phenomenon of reactive fibrosis is a primary hurdle for long term functionality of PNIs. This proof-of-concept study aimed to fabricate and test a novel, stable PNI that harnesses fibro-axonal outgrowth at the nerve end and includes fibrosis in the design. Two non-human primates were implanted with Substrate-guided, Tissue-Electrode Encapsulation and Integration (STEER) PNIs. The implant included a 3D printed guide that strove to steer the regrowing nerve towards encapsulation of the electrodes into a fibro-axonal tissue. After four months from implantation, we performed electrophysiological measurements to test STEER's functionality and examined the macro and micro- morphology of the outgrowth tissue. We observed a highly structured fibro-axonal composite within the STEER PNI. A conduction of intracranially generated action potentials was successfully recorded across the neural interface. Immunohistology demonstrated uniquely configured laminae of myelinated axons encasing the implant. STEER PNI reconfigured the structure of the fibro-axonal tissue and facilitated long-term functionality and stability of the neural interface. The results point to the feasibility of our concept for creating a stable PNI with long-term electrophysiologic functionality by using simple design and materials.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call