Abstract

Count sheets, when placed in contact with surgical instruments during steam sterilization, can transfer ink to the instruments. To explore whether this poses a safety concern, stainless steel instruments were placed on top of completely inked paper and subjected to steam sterilization, extracted, and tested for cytotoxicity. Preprinted labels were examined in a similar fashion. Extracts from stainless steel devices exposed to ink, toner, or labels showed no significant cytotoxic response, although the ink residue on the devices after steam sterilization is difficult to remove and detrimental to the instrument. Placing a barrier between the count sheet and the devices facilitates reuse of the instruments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.