Abstract
The need for better and longer lasting trileaflet valves has led to the design and development of the ABIOMED polymeric trileaflet valve prosthesis. In vitro fluid dynamic studies in the aortic position indicate that overall it has improved leaflet motion characteristics and pressure drop characteristics compared to the Carpentier-Edwards porcine and Ionescu-Shiley pericardial tissue valves in current clinical use. The ABIOMED valve is, however, more stenotic compared to the St. Jude and Medtronic-Hall low profile mechanical valves, at normal cardiac outputs. Steady and pulsatile flow velocity measurements with a laser-Doppler anemometer system indicate that the flow field downstream of the ABIOMED valve is jet-like and leads to elevated shear stresses. These shear stresses are, however, lower than those observed with the Ionescu-Shiley and Carpentier-Edwards tissue valves. The ABIOMED valves tested had been originally configured for use in valved conduits, and it is therefore our opinion that further improvements can be made to the valve and stent design which would enhance its fluid dynamic performance.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scandinavian journal of thoracic and cardiovascular surgery
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.