Abstract
The Munich pulsed low energy positron beam system (PLEPS) is now installed at the high intensity positron source (NEPOMUC) at the Munich Research Reactor FRM-II. In order to enhance the performance of the system several improvements have been implemented: two additional collinear detector ports have been installed. Therefore in addition to the normal lifetime measurements it is now possible to simultaneously perform Doppler-broadening, coincident Doppler-broadening and age momentum correlation experiments. An additional chopper was included to periodically suppress pulses and therefore to extend the standard time window of 20 ns for precise measurements of longer lifetimes. First test-experiments have been performed in May and July 2007. With all pulsing components in operation we achieved a count-rate of 1.4 × 10 4 counts per second. The total time resolution (pulsing and detector) was about 240 ps (FWHM) with a peak to background ratio up to 6 × 10 3:1.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.