Abstract

A quantitative liquid chromatographic method in which tolmetin sodium is separated from an internal standard on a C 18 column with detection at 317 nm has been developed with the aid of two statistical optimization procedures. The effects of simultaneously varying the pH, methanol-water ratio, and the concentrations of buffer and ion-pair reagent in the mobile phase were studied. A two-level factorial design was used to investigate the interactions among the variables, and the sequential simplex procedure was used to optimize the separation. A novel quality criterion was developed for the simplex optimization. Using synthetic mixtures, the mean recovery value ± S.D. ( n = 6) of tolmetin sodium was 98.7 ± 0.19% for tablets and 98.5 ± 0.12% for capsules. The assay results for commercial tablets and capsules were comparable to those obtained by the USP XXI procedure.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.