Abstract

Statistical fluctuations, or noise, in reconstructed tomograms are analyzed as a function of x-ray energy. The result is essentially constant for energies above 50 keV, assuming constant dose and spatial resolution, and is in close agreement with the actual performance of the EMI-Scanner (water-bag model). The noise is inversely proportional to the square of cell dimensions. For scanners without a water bag, the theoretical noise is approximately 30% less.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.