Abstract

The accuracy of an improved quantitative immunofluorescence method to measure the antibody binding capacity of intestinal bacteria was assessed. The improvements comprise: calibration on a non-fading fluorescence standard (uranyl glass), prolonged camera exposure time (over 4 s) and shading correction of the digitized image. This method can measure levels of IgG, IgM and IgA with variation coefficient of 6.1%, 6.4% and 9.8%, respectively. With this accuracy biological alterations in the antibody binding capacity of intestinal bacteria can be monitored at a statistically significant level.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.