Abstract

The commercial Positron Emission Tomography (PET) scanners use inorganic crystal scintillators for the detection of gamma photons. The Jagiellonian-PET (J-PET) detector exhibits high time resolution due to use of fast plastic scintillators and dedicated electronics circuits. Since the time resolution of PET scanner is influenced by numerous factors, e.g. a type of photomultipliers attached to the scintillators, the optimal selection of components of the J-PET system requires detailed understanding of the method for calculation the time resolution. In this paper we show the idea of this method, based on statistical analysis of the observed signals on the photomultiplier's output. The method is tested using signals registered by means of the single detection module of the J-PET scanner built out from 30 cm long plastic scintillator strips. We investigate two main factors affecting the photon registration probability, photomultipliers quantum efficiency and photomultipliers transit time spread. We demonstrate that the quantum efficiency of photomultipliers represents the most important factor influencing overall performance of the J-PET scanner.