Abstract

High-throughput screening (HTS) is an efficient technology for drug discovery. It allows for screening of more than 100,000 compounds a day per screen and requires effective procedures for quality control. The authors have developed a method for evaluating a background surface of an HTS assay; it can be used to correct raw HTS data. This correction is necessary to take into account systematic errors that may affect the procedure of hit selection. The described method allows one to analyze experimental HTS data and determine trends and local fluctuations of the corresponding background surfaces. For an assay with a large number of plates, the deviations of the background surface from a plane are caused by systematic errors. Their influence can be minimized by the subtraction of the systematic background from the raw data. Two experimental HTS assays from the ChemBank database are examined in this article. The systematic error present in these data was estimated and removed from them. It enabled the authors to correct the hit selection procedure for both assays.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.