Abstract

The relative strength of six different configurations of external skeletal fixation was determined by subjecting them to axial compression, and shear and torsion forces while measuring the load they would sustain before failing. The double clamp, single connecting bar, double connecting bar, quadrilateral frame, full pin splintage and three‐ dimensional tent configurations were found to be successively stronger in all three of the tests with the exception of the quadrilateral frame being more resistant to shear force than the full pin splintage configuration.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.