Abstract

The raw data acquired by Fourier-transform imaging spectrometers are the physical superposition of an interferogram and image. To reconstruct an accurate spectrum from a pure interferogram via Fourier transformation and get a pure image that is undisturbed by fringes, the interferogram and the image need to be separated. Although it can be achieved by digital image processing, heavy computations with approximation would be introduced. To overcome these drawbacks and in the meantime avoid the influence of the rapid changes of the observed scene and the perturbations of the rotating elements, a static dual-channel polarization imaging spectrometer that can simultaneously acquire inphase and antiphase interference images is presented. The extraction of a pure image and pure fringe can be simply achieved from the difference and the summation of the two interference images, respectively. The feasibility of the spectrometer and its features are described, and the influence of the polarization direction of the polarizers on the background image and fringe is discussed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call