Abstract
Sputter-ion pumps are found to start easily from nitrogen with a wet metal surface load. Repeated wet starts do not change the starting or operating characteristics of the pump. Outgassing of wet stainless steel is found to follow the relation q = 1.25 × 10 −5 t −1.47 torr-litres per second per cm 2 after 15 minutes and to be independent in shape for short or long exposures to water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.