Abstract

The possibility of the standardless mass spectrometric analysis of the elemental composition of solids has been discussed. The effect of each stage of the laser plasma expansion on the formation of the coefficient of relative sensitivity of elements in the sample has been studied theoretically and experimentally. It has been shown that the stages of ionization and detection make the main contribution to the formation of the coefficient of relative sensitivity. It has been proposed to separate dissociation and ionization processes in time and/or in space. To compensate for the energy spread of ions at the output of the analyzer, a circuit has been proposed for aligning the energy of ions before their detection.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.