Abstract

The aim of this study was to compare two immunoassay measurement methods, the ACS:180 Plus (Chiron Diagnostics) kit and the Immulite sPSA (DPC) kit. Method comparison analysis was performed according to EP9-A; approved guideline of NCCLS 51. Serum samples having a wide range of total prostate-specific antigen (PSA) concentrations were evaluated in split-sample analysis. F-test, t-test analysis and regression statistics were performed. In Deming regression analysis the coefficients were as follows; the slope=0.967; y-intercept=-0.148, r=0.989. An acceptable bias was seen since the systematical error was calculated to have a value less than the total allowable error calculated from biological variations. Non-parametric evaluation of the area under ROC curves for ACS:180 Plus and Immulite sPSA were 0.997 and 0.987, respectively. Diagnostic accuracy was at the level of p= 0.000 and no statistical difference was found between the two assay methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.