Abstract

An intercomparison on absolute activity measurements of125I sources by the sumpeak method was carried out in 1993 between the Primary Standard Laboratories for Ionizing Radiations of Italy (ENEA) and Albania (INP). The sources were prepared by ENEA starting from a125I solution. The sources were measured at ENEA and successively at INP. The results were in good agreement, the deviation of the INP results from the ENEA results being about −0.5% with an overall uncertainty of both results about 0.3% (1σ). This paper describes the instrumentation and the experimental procedures used in this comparison by the two Institutes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.