Abstract

A twin type thermal conduction calorimeter has been specially designed and constructed for absolute measurement of tritium, and then its performance was examined using two kinds of tritium samples, i.e., ZrNi alloy powders loaded with a given amount of tritium. Total amounts of tritium in the ZrNi alloy powders were previously estimated by a small ionization chamber. Changes in radioactivity of two samples were followed for three years, and it was seen that the activity changes obeys the half-life of 4505 days. The initial tritium amounts estimated by the small ionization chamber agreed well with the extrapolated values from the activity changes evaluated by the calorimeter. Namely, it was suggested that the small ionization chamber is able to be used as a secondary standard of the tritium measuring devices. The β-ray-induced X-ray counter was also calibrated for non-destructive measurements of high-level tritium by using the present small ionization chamber.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.