Abstract

AbstractThe spectral sensitivities of different energy‐dispersive X‐ray fluorescence spectrometers may vary to the extent that the same glass fragment produces noticeably different spectra. In this study a set of twelve glass standards has been prepared. These are polished glass discs which span a wide range of elemental levels, and have been prepared for the monitoring and measurement of relative instrument sensitivies. The polished discs were analysed several times in rapid succession on one instrument and the mean elemental peak area ratios obtained were designated as standard values for these samples. During subsequent routine analytical work on this and other instruments, the discs are regularly analysed and if any significant changes are found the new elemental ratios can be corrected to the standard values usually simply by multiplying by a constant (the relative sensitivity factor). The procedure is shown to be effective in the case of 14 fragments of glass weighing between 40 and 1160 μg which were analysed several months apart on two different instruments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.