Abstract

Standardization of 153Sm by 4π(LS)-γ coincidence and anticoincidence counting and the CIEMAT/NIST method in three LS-counters is presented. This short half-life radionuclide is applied in tumor therapy and bone pain palliation. A simplified disintegration scheme of 153Sm was applied in the calculation of the counting efficiency. Standard uncertainties of 0.4% for the 153Sm measurements by the 4π(LS)-γ coincidence and anticoincidence techniques and 0.7% by the C/N method were evaluated, respectively. An agreement of the standardization results by both methods within the respective uncertainties was obtained. The half-life of 153Sm of (1.92895±0.00024) days was determined during one month of measurements and correction for europium isotope impurities by the C/N method in the TriCarb 2910 LS-counter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.