Abstract

A solution of 124I was standardized for activity by 4πβ(LS)-γ(NaI) live-timed anticoincidence (LTAC) counting, with confirmatory measurements by triple-to-double coincidence ratio (TDCR) and CIEMAT-NIST efficiency tracing (CNET) liquid scintillation counting. The LTAC-based standard was shown to be in agreement (within k = 1 uncertainties) with previous measurements at NIST and elsewhere. Calibration settings for radionuclide calibrators were determined and a discrepancy with literature values, partially due to a calibration methodology dependent upon an erroneous setting for 18F, was identified and explained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.