Abstract

The procedure followed by the Nuclear Metrology Laboratory, at IPEN, for the primary standardization of 99mTc is described. The primary standardization has been accomplished by the coincidence method. The beta channel efficiency was varied by electronic discrimination using a software coincidence counting system. Two windows were selected for the gamma channel: one at 140keV gamma-ray and the other at 20keV X-ray total absorption peaks. The experimental extrapolation curves were compared with Monte Carlo simulations by means of code ESQUEMA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.