Abstract
From a commercial supplier, three independent lots of an aqueous solution containing 223Ra in equilibrium with its deteriorating progeny were considered for standardization, by using live-time anticoincidence counting (LTAC) in the Laboratório Nacional de Metrologia das Radiações Ionizantes (LNMRI) Brazil. The ionization chamber calibration factors were obtained using measurements of independent lots of 223Ra in LTAC considering the absolute method in order to evaluate the constancy and reproducibility of the standardization. The calibration factors of three high-performance well-type ionization chamber systems and four commercial well-type ionization chambers were determined. The LTAC results showed that lots 2 and 3 were 1069.88 and 1097.44 kBq/kg, with corresponding relative standard uncertainties 0.12% and 0.20%, respectively. The spectrometry method results of each lot value were classified as positive by ANOVA hypothesis testing. The emission probabilities relating to the X- and γ-rays in region from 81 to 830 keV ware determined.Hence, LNMRI can provide calibration services for 223Ra.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.