Abstract

A simple distance-variable detection table on which two detectors can be mounted and moved back and forth facing each other centred on a sample source has been constructed for photon–photon and α–photon coincidence counting for 125I and 238Pu, respectively. Two NaI(Tl) detectors were used for the standardization of 125I. A Si detector and an NaI(Tl) detector were employed for the standardization of 238Pu. The NaI(Tl) detectors have a very thin aluminium window for measuring low-energy photons from 125I and 238Pu while the Si-detector is a usable-in-air ion-implanted detector for α particles from 238Pu. A thin collodion film was used for 238Pu samples, whereas a thick polyester film was used for 125I samples to prevent 125I from leaking from the sample. The key comparison result of 125I showed good agreement, and the result of 238Pu was also compared with that of liquid scintillation counting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.