Abstract
The nuclide 110mAg is a β–γ emitter with a very complex decay scheme, including more than 50 γ-rays. It has been standardized by two methods: liquid scintillation counting (LSC) and 4πβ(pc)-γ coincidence measurements. In the LSC measurements the CIEMAT/NIST method was used, with 3H being used as a tracer for efficiency calculations. In the 4πβ(pc)-γ standardization, Monte Carlo calculations have been made to determine the optimal measurement conditions. Results obtained with both methods for the activity concentration of the solution are in good agreement.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.