Abstract

The use of Rhenium-188 for various therapeutic applications in the field of nuclear medicine has increased in recent years due to its favourable properties like decay scheme, cost effective availability and easy chemistry. Two independent measuring setups were used to standardise 188Re radioactive solution. The modus operandi of standardisation was 4πβ-γ coincidence technique where the beta detection was done by proportional counting and liquid scintillation counting and the gamma detection was done by using NaI(Tl) detectors. The secondary standard, high pressure ionisation chamber type Centronic IG12, 20A was calibrated with the standardised 188Re solution and the sensitivity coefficient (pA MBq−1) was determined. To enhance the accuracy of the commercial radionuclide calibrator and to ensure that patients receive optimum dose of these radiopharmaceuticals, calibration number of the Capintec CRC-15β radionuclide calibrator was also verified. This paper presents the standardisation of 188Re radioactive solution by primary methods and calibration of BARC secondary standard ionisation chamber system and a Capintec CRC-15β radionuclide calibrator.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.