Abstract

This paper presents the work performed in IFIN-HH, Radionuclide Metrology Laboratory, for the standardization of a 68(Ge+Ga) solution, within our participation in the key-comparison codified as CCRI(II)-K2.Ge-68, organised by the CIPM-CCRI(II). Four methods were used: (i) the 4πβ(PC)-γ coincidence method, efficiency extrapolation variant, (ii) the liquid scintillation counting in the variant of triple to double coincidence ratio, (iii) the gamma-ray spectrometry and (iv) the well-type ionisation chamber measurement. The final result was calculated as “power moderate weighted mean” of all results. The results obtained in all methods are presented as relative figures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.