Abstract

The standardisation of 111In by 4π(LS)-γ coincidence and anticoincidence counting is presented. Absolute measurements were performed for samples with different concentrations of carrier solution and for different window settings in the gamma channel. The radioactive concentration of the master solution determined on the same reference date was consistent for all measurements performed. The evaluated typical uncertainty was 0.43%. The half-life of 111In was determined using a time series of measurements performed with an ionisation chamber. A least squares fit of the measured data resulted in a half-life of 2.8067 (34) days consistent with Decay Data Evaluation Project recommended value (0.064% higher than the DDEP value).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.