Abstract

Ultra-short pulse laser systems are considered as promising tools in the fields of precise micro-machining and medicine applications. In the course of the development of reliable table top laser systems, a rapid growth of ultra-short pulse applications could be observed during the recent years. The key for improving the performance of high power laser systems is the quality of the optical components concerning spectral characteristics, optical losses and the power handling capability. In the field of ultra-short pulses, standard measurement procedures in quality management have to be validated in respect to effects induced by the extremely high peak power densities. The present work, which is embedded in the EUREKA-project CHOCLAB II, is predominantly concentrated on measuring the multiple-pulse LIDT (ISO 11254-2) in the fs-regime. A measurement facility based on a Ti:Sapphire-CPA system was developed to investigate the damage behavior of optical components. The set-up was supplied with an improved pulse energy detector discriminating the influence of pulse-to-pulse energy fluctuations on the incidence of damage. Aditionally, a laser-calorimetric measurement facility determining the absorption (ISO 11551) utilizing a fs-Ti:Sapphire laser was accomplished. The investigation for different pulse durations between 130 fs and 1 ps revealed a drastic increase of absorption in titania coatings for ultra-short pulses.